Z-1329-06
Pointe Scientific Liquid Glucose (HEXO) Reagent Set, for the quantitative determination of glucose in serum...
Horiba Instruments Incorporated / initiated 2006-06-07 / Terminated / root cause: labeling and UDI
Pointe Scientific, Inc. began a recall of Pointe Scientific Liquid Glucose (HEXO) Reagent Set, for the quantitative determination of glucose in serum, 10 x 100 ml R1, 10 x 20 ml R2 glucose reagent, or... on . The reason FDA records is "Product does not meet performance specifications through its labeled expiration period". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1329-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Horiba Instruments Incorporated
- Product
- Pointe Scientific Liquid Glucose (HEXO) Reagent Set, for the quantitative determination of glucose in serum, 10 x 100 ml R1, 10 x 20 ml R2 glucose reagent, or 1 x 2 L R1 x 500 ml R2, for use with Hitachi analyzer; in vitro diagnostic, Catalog numbers HG920-1200 and HG920-2500.
- Product code
- CFR Hexokinase, Glucose
- Reason for recall
- Product does not meet performance specifications through its labeled expiration period.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Customers were notified via recall letter dated 6/7/06 to cease using and to discard these lots, and to send back the recall response form for replacement stocks.
- Quantity in commerce
- 153 kits
- Distribution
- Worldwide, including USA, Canada, China, Panama, Russia and Trinidad.
- Product codes and lots
- Lots 431704, exp. 11/06 and 511901, exp. 4/07.