Z-1230-06

Dyonics Electroblade, 4.5 Full Radius, Valleylab Generator-Compatible Part Number: 7205961

Smith & Nephew, Inc. / initiated 2006-06-19 / Terminated / root cause: packaging

Smith & Nephew, Inc. Endoscopy Division began a recall of Dyonics Electroblade, 4.5 Full Radius, Valleylab Generator-Compatible Part Number: 7205961 on . The reason FDA records is "Sterility seal of the package tray may have gaps/voids compromising the sterility of the device". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1230-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Dyonics Electroblade, 4.5 Full Radius, Valleylab Generator-Compatible Part Number: 7205961
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Sterility seal of the package tray may have gaps/voids compromising the sterility of the device.
Root cause tag
packaging
FDA root cause
Other
Action
Smith & Nephew notified consignees by Certified letter dated 6/19/06. Accounts are requested to return product. Smith & Nephew expanded the recall on 8/2/5/06 by letter to include addtional lot numbers.
Quantity in commerce
445 boxes/Expanded 348 units
Distribution
Nationwide Foreign: Smith & Nephew Canada, Germany, France, Istanbul
Product codes and lots
Lot Numbers: Equal to or lower 50151697 Expanded Recall: Lot Numbers: 50173181, 50175084