Z-1405-06

Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog No. 801046

Terumo Cardiovascular Systems Corporation / initiated 2006-06-30 / Terminated

Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog No. 801046 on . The reason FDA records is "The roller pump speed control knobs and the centrifugal pump speed control knobs may fail to function/change pump speed". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1405-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog No. 801046.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
The roller pump speed control knobs and the centrifugal pump speed control knobs may fail to function/change pump speed.
FDA root cause
Other
Action
The firm replaced the graphics driver board on complaint units as user complaints were received. The firm notified all users of the problem and the presence of an alternative method of pump speed control via recall letter dated 6/30/06.
Quantity in commerce
Recall Numbers Z-1403-1405-2006 -1,135 systems in the U.S. plus 723 internationally.
Distribution
Worldwide-USA, Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Product codes and lots
Serial numbers 0020 through 0216.