Z-1405-06
Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog No. 801046
Terumo Cardiovascular Systems Corporation / initiated 2006-06-30 / Terminated
Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog No. 801046 on . The reason FDA records is "The roller pump speed control knobs and the centrifugal pump speed control knobs may fail to function/change pump speed". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1405-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog No. 801046.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- The roller pump speed control knobs and the centrifugal pump speed control knobs may fail to function/change pump speed.
- FDA root cause
- Other
- Action
- The firm replaced the graphics driver board on complaint units as user complaints were received. The firm notified all users of the problem and the presence of an alternative method of pump speed control via recall letter dated 6/30/06.
- Quantity in commerce
- Recall Numbers Z-1403-1405-2006 -1,135 systems in the U.S. plus 723 internationally.
- Distribution
- Worldwide-USA, Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
- Product codes and lots
- Serial numbers 0020 through 0216.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
