Z-1413-06
Disetronic D-TRONplus Insulin Pump Battery Pack; Ref/Catalog no. 04697014001
Disetronic Medical Systems / initiated 2006-07-13 / Terminated / root cause: design
Disetronic Medical Systems, Inc. began a recall of Disetronic D-TRONplus Insulin Pump Battery Pack; Ref/Catalog no. 04697014001 on . The reason FDA records is "The battery may turn the pump off without warning due to a design change in the battery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1413-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Disetronic Medical Systems
- Product
- Disetronic D-TRONplus Insulin Pump Battery Pack; Ref/Catalog no. 04697014001.
- Product code
- LZG Pump, Infusion, Insulin
- Reason for recall
- The battery may turn the pump off without warning due to a design change in the battery.
- Root cause tag
- design
- FDA root cause
- Other
- Action
- Press release was issued on 7/13/06. Consignees, distributors and health care providers were issued a recall notice dated 7/13/06 instructing replacement of the power pack every two weeks until further notice.
- Quantity in commerce
- 6,604 packs to U.S. and Canada.
- Distribution
- Nationwide and Canada.
- Product codes and lots
- Lots 06010107, 06030242, 06040085, 06050110 and 06060126.