Package Insert which accompanies the Hematocrit and Metabolite QUALICHECK LEVEL 2 Quality Control of...
Radiometer America Inc began a recall of Package Insert which accompanies the Hematocrit and Metabolite QUALICHECK LEVEL 2 Quality Control of Hematocrits and Metabolites, used with the ABL 77/70/555... on . The reason FDA records is "The package insert, which accompanied the QC reagents used with ABL blood gas analyzers, referenced incorrect control ranges for Hct (hematocrit) testing. The range stated on the package insert was too low to generate...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1316-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Radiometer America, Inc
- Product
- Package Insert which accompanies the Hematocrit and Metabolite QUALICHECK LEVEL 2 Quality Control of Hematocrits and Metabolites, used with the ABL 77/70/555 Blood Gas & Electrolyte Analyzers. The QC reagents are packaged in a box of 30 glass vials of (1.5 ml) per box. The package insert is packed one per box.
- Product code
- GKF Instrument, Hematocrit, Automated
- Reason for recall
- The package insert, which accompanied the QC reagents used with ABL blood gas analyzers, referenced incorrect control ranges for Hct (hematocrit) testing. The range stated on the package insert was too low to generate accurate Hct test results (The control ranges were acceptable for the glucose and lactate testing).
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- The firm apprised their customers of this situation with a letter dated 5/24/2005. The letter requests that the customer remove the old (incorrect) package insert and replace it with a new (correct) insert which accompanied the customer letter. Should the customer have any questions, or need additional package inserts, the letter provides a toll-free contact telephone number.
- Quantity in commerce
- 55 boxes x 1 package insert per box
- Distribution
- Nationwide, including: PA, MI, NY, OH, NJ, TX, AL, MO, AR, TN, WA, KY, and NC, and London, ON, Canada
- Product codes and lots
- Model #.S7180, Part Number: 944040, Lot #15.