Z-1466-06
FP1000 Cell Preparation System Part Number 624922
Siemens Healthcare Diagnostics Inc. / initiated 2006-06-26 / Terminated
Beckman Coulter Inc began a recall of FP1000 Cell Preparation System Part Number 624922 on . The reason FDA records is "During the cleaning cycle performed during the shutdown procedure of the Beckman Coulter FP1000 Cell Preparation System, fluid (diluted bleach) may drip from the probe in the location of the back reagent rack...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1466-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- FP1000 Cell Preparation System Part Number 624922
- Product code
- LXG Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use
- Reason for recall
- During the cleaning cycle performed during the shutdown procedure of the Beckman Coulter FP1000 Cell Preparation System, fluid (diluted bleach) may drip from the probe in the location of the back reagent rack potentially resulting in bleach and/or water dripping into the reagents contaminating them and resulting in possible incorrect results.
- FDA root cause
- Other
- Action
- A Product Corrective Action letter was sent on June 26, 2006 informing users that during the cleaning cycle performed during the shutdown procedure of the Beckman Coulter FP1000 Preparation System, fluid (diluted bleach) may drip from the probe in the location of the back reagent rack potentially resulting in bleach and/or water dripping into the reagents contaminating them and resulting in possible incorrect results. To reduce the risk of incorrect results from dripping fluid, users were advised to remove all patient samples, reagents, controls, and calibrators before initiating the shutdown cycle of the Beckman Coulter FP1000. Software will be modified in the next version to prevent the possibility of dripping bleach during the shutdown cycle. The Product Corrective Action letter is to be maintained as part of the customer''s laboratory Quality System documentation. The enclosed Response Form is to be completed and returned within 10 days to the recalling firm. Customers are also advised to contact the firm or their local Beckman Coulter Representative if there are questions.
- Quantity in commerce
- 11 units distributed
- Distribution
- Nationwide and Canada
- Product codes and lots
- Software Version 1.0