GammaMed software program, version 5.07, vers. 5.08, vers.-usa, vers.-tst and vers.-row , part number...
Varian Medical Systems began a recall of GammaMed software program, version 5.07, vers. 5.08, vers.-usa, vers.-tst and vers.-row , part number GM11005400, for the GammaMed model 12it radionuclide... on . The reason FDA records is "Software control program for a medical device used in radiation treatment may cause practitioners to incorrectly administer the treatment plan to cancer patients. There is a hazard when entering the treatment plans...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1399-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- GammaMed software program, version 5.07, vers. 5.08, vers.-usa, vers.-tst and vers.-row , part number GM11005400, for the GammaMed model 12it radionuclide applicator system.
- Product code
- JAQ System, Applicator, Radionuclide, Remote-Controlled
- Reason for recall
- Software control program for a medical device used in radiation treatment may cause practitioners to incorrectly administer the treatment plan to cancer patients. There is a hazard when entering the treatment plans manually that the user neglects to change the default step size and/or the origin or accidentally enters incorrect parameters.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- The recalling firm notified healthcare practitioners using the medical device by letter on 4/17/06 and continuing through 5/1/06. The notification was flagged as a 'Medical Device Recall' and instructed users to adjust system parameters in the autoexec.bat file. The notification further recommended using the identified treatment plans and that manual input should be used for QA purposes only.
- Quantity in commerce
- 13 units
- Distribution
- Nationwide.
- Product codes and lots
- GammaMed ''12it'' serial numbers: GM00213 GM00218 GM00219TB GM00215 GM00202 GM00216 GM00217 GM00212 GM00219TA GM00219 GM00210 GM00214 GM0213