Z-1436-06
D826 Accessory kit for single tcpO2 Electrodes for use with the TCM400 Transcutaneous Monitor (Cutaneous...
Radiometer America, Inc / initiated 2005-11-01 / Terminated / root cause: manufacturing process
Radiometer America Inc began a recall of D826 Accessory kit for single tcpO2 Electrodes for use with the TCM400 Transcutaneous Monitor (Cutaneous Oxygen Monitor) on . The reason FDA records is "Excessive Drift-the membrane units of the device cause the electrode to exceed performance standards for drift in the first two or more calibrations". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1436-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Radiometer America, Inc
- Product
- D826 Accessory kit for single tcpO2 Electrodes for use with the TCM400 Transcutaneous Monitor (Cutaneous Oxygen Monitor)
- Product code
- KLK Monitor, Oxygen, Cutaneous, For Infant Not Under Gas Anesthesia
- Reason for recall
- Excessive Drift-the membrane units of the device cause the electrode to exceed performance standards for drift in the first two or more calibrations
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- The firm decided to initiate a recall on 9/8/2005 and the recall letter was issued in November, 2005. The letter informs the customers of this problem and asks that each customer inspect their current stock for the presence of the recalled lot codes and discard any such product found. The letter is accompanied by a recall response form which the customer can use to request additional replacement ellectrodes as necessary.
- Quantity in commerce
- 3,323-USA, 1,197-Foreign
- Distribution
- Worldwide distribution --- including the states of PA, IL, TX, NY, FL, KS, NE, AL, MI, LA, WI, MN, NJ, CA, NH, MD, OH, MO, UT, IN, GA, WA, MA, KY, NC, IA, OK, SD, CO, AR, SC, ME, NV, AZ, RI, AK, MS, VA, WV, MT, TN, IN & NM. Country List pending
- Product codes and lots
- Type D826, Accessory Kit Part Number: 904-308. Lot Codes R0172 through R0223, inclusive.