Roche Tina-Quant ASLO (Anti-streptolysin O) Reagent Kit; Catalog No. 1931601 (Roche Material No. 11931601216)...
Roche Diagnostics Corp. began a recall of Roche Tina-Quant ASLO (Anti-streptolysin O) Reagent Kit; Catalog No. 1931601 (Roche Material No. 11931601216) Lot No. 674653 on . The reason FDA records is "The vial containing the R2 reagent is labeled as R1 reagent and if placed in the R1 slot of the analyzer, erroneous results will be obtained. (extremely low patient and control results)". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1385-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche Tina-Quant ASLO (Anti-streptolysin O) Reagent Kit; Catalog No. 1931601 (Roche Material No. 11931601216) Lot No. 674653
- Product code
- GTQ Antistreptolysin - Titer/Streptolysin O Reagent
- Reason for recall
- The vial containing the R2 reagent is labeled as R1 reagent and if placed in the R1 slot of the analyzer, erroneous results will be obtained. (extremely low patient and control results)
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Consignees were notified via recall letter dated 7/25/06 to cease using and to discard the product.
- Quantity in commerce
- 20 kits.
- Distribution
- Nationwide including the states of Arizona, California, New Jersey and Pennsylvania.
- Product codes and lots
- Lot 674653; exp. 12/31/2006.