Z-1385-06

Roche Tina-Quant ASLO (Anti-streptolysin O) Reagent Kit; Catalog No. 1931601 (Roche Material No. 11931601216)...

Roche Diagnostics GmbH / initiated 2006-07-25 / Terminated / root cause: labeling and UDI

Roche Diagnostics Corp. began a recall of Roche Tina-Quant ASLO (Anti-streptolysin O) Reagent Kit; Catalog No. 1931601 (Roche Material No. 11931601216) Lot No. 674653 on . The reason FDA records is "The vial containing the R2 reagent is labeled as R1 reagent and if placed in the R1 slot of the analyzer, erroneous results will be obtained. (extremely low patient and control results)". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1385-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche Tina-Quant ASLO (Anti-streptolysin O) Reagent Kit; Catalog No. 1931601 (Roche Material No. 11931601216) Lot No. 674653
Product code
GTQ Antistreptolysin - Titer/Streptolysin O Reagent
Reason for recall
The vial containing the R2 reagent is labeled as R1 reagent and if placed in the R1 slot of the analyzer, erroneous results will be obtained. (extremely low patient and control results)
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Consignees were notified via recall letter dated 7/25/06 to cease using and to discard the product.
Quantity in commerce
20 kits.
Distribution
Nationwide including the states of Arizona, California, New Jersey and Pennsylvania.
Product codes and lots
Lot 674653; exp. 12/31/2006.