Z-0016-2007
Joerns Ultra Care 770 Model AC Powered Beds with model numbers U770, U770 AL , U770 GNDAL. Sunrise Medical...
Joerns Healthcare / initiated 2006-01-01 / Terminated
Sunrise Medical CCG, Inc. began a recall of Joerns Ultra Care 770 Model AC Powered Beds with model numbers U770, U770 AL , U770 GNDAL. Sunrise Medical, Long Term Care Division, 5001 Joerns Drive, Stevens... on . The reason FDA records is "The control box used in the actuator systems appears to have a potential to become stuck in one position resulting a non-functioning bed". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0016-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Joerns Healthcare
- Product
- Joerns Ultra Care 770 Model AC Powered Beds with model numbers U770, U770 AL , U770 GNDAL. Sunrise Medical, Long Term Care Division, 5001 Joerns Drive, Stevens Point, WI 54481 USA.
- Product code
- FNL Bed, Ac-Powered Adjustable Hospital
- Reason for recall
- The control box used in the actuator systems appears to have a potential to become stuck in one position resulting a non-functioning bed.
- FDA root cause
- Other
- Action
- The objective is to replace all of these systems. Phone contact of affected facilities began in January 2006. All facilities to be contacted by June 23, 2006. All products will be corrected at the location of the end-user by June 30, 2006.
- Quantity in commerce
- 6,921
- Distribution
- Nationwide and Canada.
- Product codes and lots
- Model U 770, serial number range: 6149310001 - 6149310005, 6151360001 - 6151360005, 6163290001 - 616330001. Model U770 AL, serial number range: 6099280001 - 6113410042, 6114760001 - 6127920006, 6128720001 - 6130360001, 6130510001 - 6135430002, 6136650001 - 6140860005, 6142190001 - 6142320001, 6142380001 - 6148440006, 6149320001 - 6151270004, 6152090001 - 6160890005, 6161480001 - 6162420010, 6163330001 - 6168440002, 6169170001 - 6171470001, 6171750001 - 6174080020, 6174750001 - 6177310002. U770-GNDAL, serial number range: 6113510001 - 6113510004, 6127970001 - 61279700015, 6130370001 - 6130370003, 6135440001, 6140870001 - 61408700010, 6142330001, 6160900001 - 61609000010, 6168450001 - 6168450004, 6171700001 - 6171720002, 6174710001 - 61747100012.