Z-1324-06

Reference Electrodes for ABL 700/800 Blood Gas Analyzers. Electrode Part Numbers: 945-633, distributed one...

Radiometer America, Inc / initiated 2005-09-19 / Terminated

Radiometer America Inc began a recall of Reference Electrodes for ABL 700/800 Blood Gas Analyzers. Electrode Part Numbers: 945-633, distributed one (1) electrode per box on . The reason FDA records is "Reference electrodes for the firm's ABL 700/800 Blood Gas Analyzers are oversized and will not fit into the instrument". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1324-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Radiometer America, Inc
Product
Reference Electrodes for ABL 700/800 Blood Gas Analyzers. Electrode Part Numbers: 945-633, distributed one (1) electrode per box.
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
Reference electrodes for the firm's ABL 700/800 Blood Gas Analyzers are oversized and will not fit into the instrument.
FDA root cause
Other
Action
Following the receipt of three (3) customer complaints, on 9/14/2005, the firm issued an E-Mail notification [Management Cover] to their Service Personnel on 9/19/2005. The Management Cover instructed the service personnel to check the electrodes located at their stock and their customer locations; put all electrodes, which cannot be inserted in their call-back stock; and provide for credit to their customers through normal channels. The Service Personnel in turn, contacted their customers via telephone on or around 9/20/2005 and asked that they check their inventories for the presence of any electrodes which would not fit in the analyzer and return any such electrodes to Radiometer America for replacement
Quantity in commerce
9
Distribution
Nationwide and Canada.
Product codes and lots
Part # 945-633 - Lot #96-02 and Lot #96-03