Z-1277-06
Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 14 mm diameter, 130 mm length, non-sterile...
Zimmer GmbH / initiated 2006-06-16 / Terminated
Zimmer Inc. began a recall of Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 14 mm diameter, 130 mm length, non-sterile, catalog no. (00-4309-014-13) on . The reason FDA records is "During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1277-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 14 mm diameter, 130 mm length, non-sterile, catalog no. 4309-14-13 (00-4309-014-13).
- Product code
- HWT Template
- Reason for recall
- During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- FDA root cause
- Other
- Action
- U.S. and international distributors were notified of the recall via letter dated 6/30/2006 to check customer inventories and to return recalled product to the firm. Trial implants were redesigned and reworked to eliminate the problem.
- Quantity in commerce
- 324 total
- Distribution
- Nationwide and worldwide including Australia, Canada, Germany, Italy and Spain.
- Product codes and lots
- All lots.
FIRST ARTICLE / WEEKLY