Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1279-06

Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 16 mm diameter, 130 mm length, non-sterile...

Zimmer GmbH / initiated 2006-06-16 / Terminated

Zimmer Inc. began a recall of Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 16 mm diameter, 130 mm length, non-sterile, catalog no. (00-4309-016-13) on . The reason FDA records is "During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1279-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 16 mm diameter, 130 mm length, non-sterile, catalog no. 4309-16-13 (00-4309-016-13).
Product code
HWT Template
Reason for recall
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
FDA root cause
Other
Action
U.S. and international distributors were notified of the recall via letter dated 6/30/2006 to check customer inventories and to return recalled product to the firm. Trial implants were redesigned and reworked to eliminate the problem.
Quantity in commerce
324 total
Distribution
Nationwide and worldwide including Australia, Canada, Germany, Italy and Spain.
Product codes and lots
All lots.