Z-1341-06
Zimmer M/DN Intramedullary Fixation 4.2 dia. cortical screw, 3.5 mm hex head, 85 mm length, 22-13-5 stainless...
Zimmer GmbH / initiated 2006-06-26 / Terminated / root cause: packaging
Zimmer Inc. began a recall of Zimmer M/DN Intramedullary Fixation 4.2 dia. cortical screw, 3.5 mm hex head, 85 mm length, 22-13-5 stainless steel, sterile bone screw; Cat. no. ... on . The reason FDA records is "Lack of assurance of sterility, as the sterile barrier packaging may have been compromised". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1341-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer M/DN Intramedullary Fixation 4.2 dia. cortical screw, 3.5 mm hex head, 85 mm length, 22-13-5 stainless steel, sterile bone screw; Cat. no. 2253-85-42 (00225308542).
- Product code
- NDJ Screw, Fixation, Bone, Non-Spinal, Metallic
- Reason for recall
- Lack of assurance of sterility, as the sterile barrier packaging may have been compromised.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Hospital accounts were notified via recall letter dated 6/26/06 and distributor accounts were notified via recall letter dated 6/28/06 to return the product. Recall was extended to additional lots via recall letter dated 7/26/06.
- Quantity in commerce
- 310 total for all catalog numbers
- Distribution
- Worldwide, including USA, Argentina, Brazil, Canada, China, England, France, Germany, Guatemala, Israel, Japan, Jordan, Lebanon, Poland, Saudi Arabia, Singapore, Spain, Taiwan, UAE.
- Product codes and lots
- Lots 15887800, 16568000, 17310000, 17310100, 17657500, 17734300 and 23083300.