Z-1342-06

Zimmer M/DN Intramedullary Fixation 5.5 dia. cortical screw, 3.5 mm hex head, 95 mm length, 22-13-5 stainless...

Zimmer GmbH / initiated 2006-06-26 / Terminated / root cause: packaging

Zimmer Inc. began a recall of Zimmer M/DN Intramedullary Fixation 5.5 dia. cortical screw, 3.5 mm hex head, 95 mm length, 22-13-5 stainless steel, sterile bone screw; Cat. no. ... on . The reason FDA records is "Lack of assurance of sterility, as the sterile barrier packaging may have been compromised". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1342-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer M/DN Intramedullary Fixation 5.5 dia. cortical screw, 3.5 mm hex head, 95 mm length, 22-13-5 stainless steel, sterile bone screw; Cat. no. 2253-95-55 (00225309555).
Product code
NDJ Screw, Fixation, Bone, Non-Spinal, Metallic
Reason for recall
Lack of assurance of sterility, as the sterile barrier packaging may have been compromised.
Root cause tag
packaging
FDA root cause
Other
Action
Hospital accounts were notified via recall letter dated 6/26/06 and distributor accounts were notified via recall letter dated 6/28/06 to return the product. Recall was extended to additional lots via recall letter dated 7/26/06.
Quantity in commerce
310 total for all catalog numbers
Distribution
Worldwide, including USA, Argentina, Brazil, Canada, China, England, France, Germany, Guatemala, Israel, Japan, Jordan, Lebanon, Poland, Saudi Arabia, Singapore, Spain, Taiwan, UAE.
Product codes and lots
Lots 17036400, 17658000, 17734800, 17887700, 18279700, 21846200, 22983000 and 23083900.