Z-1547-06

Alcohol Control, Level 1 sold by Pointe Scientific; Catalog # 7-A7504-CTL-L1

Horiba Instruments Incorporated / initiated 2006-06-09 / Terminated

Pointe Scientific, Inc. began a recall of Alcohol Control, Level 1 sold by Pointe Scientific; Catalog # 7-A7504-CTL-L1 on . The reason FDA records is "Open vial stability-Potential microbial growth in the alcohol standard and possible unexpected QC changes in the ammonia portion of the control due to the lack of validation for long-term open vial stability". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1547-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
Alcohol Control, Level 1 sold by Pointe Scientific; Catalog # 7-A7504-CTL-L1.
Product code
JIX Calibrator, Multi-Analyte Mixture
Reason for recall
Open vial stability-Potential microbial growth in the alcohol standard and possible unexpected QC changes in the ammonia portion of the control due to the lack of validation for long-term open vial stability.
FDA root cause
Other
Action
Consignees were notified via recall letter dated 6/9/06 and a second letter dated 6/21/06 extending the recall to additional products/lots, to cease using the products, conduct a recall to the user level, destroy products on hand and to notify the firm via the response form of the amount destroyed for replacement purposes.
Quantity in commerce
98 units
Distribution
Worldwide distribution ---- including USA and countries of Canada, Ecuador, Germany, Greece, India, Mexico, Peru, Poland, Portugal, Trinidad, United Arab Emirates and United Kingdom.
Product codes and lots
Lot 522202, exp. 7/06.