Z-1549-06
Alcohol Standard sold by Pointe Scientific; Catalog # 7-A7504-STD
Horiba Instruments Incorporated / initiated 2006-06-09 / Terminated
Pointe Scientific, Inc. began a recall of Alcohol Standard sold by Pointe Scientific; Catalog # 7-A7504-STD on . The reason FDA records is "Open vial stability-Potential microbial growth in the alcohol standard and possible unexpected QC changes in the ammonia portion of the control due to the lack of validation for long-term open vial stability". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1549-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Horiba Instruments Incorporated
- Product
- Alcohol Standard sold by Pointe Scientific; Catalog # 7-A7504-STD.
- Product code
- JIX Calibrator, Multi-Analyte Mixture
- Reason for recall
- Open vial stability-Potential microbial growth in the alcohol standard and possible unexpected QC changes in the ammonia portion of the control due to the lack of validation for long-term open vial stability.
- FDA root cause
- Other
- Action
- Consignees were notified via recall letter dated 6/9/06 and a second letter dated 6/21/06 extending the recall to additional products/lots, to cease using the products, conduct a recall to the user level, destroy products on hand and to notify the firm via the response form of the amount destroyed for replacement purposes.
- Quantity in commerce
- 44 units
- Distribution
- Worldwide distribution ---- including USA and countries of Canada, Ecuador, Germany, Greece, India, Mexico, Peru, Poland, Portugal, Trinidad, United Arab Emirates and United Kingdom.
- Product codes and lots
- Lot 509102; exp. 9/06.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
