Roche MagNA Pure LC Instrument Kit; Catalog no. 12236931001. An automated, general purpose instrument for...
Roche Diagnostics Corp. began a recall of Roche MagNA Pure LC Instrument Kit; Catalog no. 12236931001. An automated, general purpose instrument for isolation and purification of nucleic acids and... on . The reason FDA records is "If the user opens the instrument lid during decontamination, the cycle will stop but the timer will continue resulting in the user believing the instrument has been decontaminated when it has not been". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1373-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche MagNA Pure LC Instrument Kit; Catalog no. 12236931001. An automated, general purpose instrument for isolation and purification of nucleic acids and assembly of PCR reactions.
- Product code
- NPQ Test, Factor V Leiden Mutations, Genomic Dna Pcr
- Reason for recall
- If the user opens the instrument lid during decontamination, the cycle will stop but the timer will continue resulting in the user believing the instrument has been decontaminated when it has not been.
- FDA root cause
- Other
- Action
- Consignees were notified via recall letter dated 7/17/06 to restart the sterilization cycle if they open the instrument for any reason during the sterilization cycle.
- Quantity in commerce
- 771 total for recalls Z-1372-1374-06
- Distribution
- Nationwide.
- Product codes and lots
- All units.