Z-1431-06
Mallinckrodt Satin-Slip Intubation Stylet, 6 Fr., single use, Model number 85863,shipped in cases of 20
Medtronic, Inc. / initiated 2006-07-12 / Terminated
Nellcor Puritan Bennett began a recall of Mallinckrodt Satin-Slip Intubation Stylet, 6 Fr., single use, Model number 85863,shipped in cases of 20 on . The reason FDA records is "Product's outer sheath material may separate from the wire interior during use and may lodge in the endotracheal tube or enter the patient's airway and interfere with ventilation". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1431-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Mallinckrodt Satin-Slip Intubation Stylet, 6 Fr., single use, Model number 85863,shipped in cases of 20
- Product code
- BSR Stylet, Tracheal Tube
- Reason for recall
- Product's outer sheath material may separate from the wire interior during use and may lodge in the endotracheal tube or enter the patient's airway and interfere with ventilation.
- FDA root cause
- Other
- Action
- The firm has issued notification letters via USPS certified mail with return receipt, requesting that the recipients who have further distributed the product forward a copy of that letter. The firm intends to contact non-responding companies beginning 30 days after the initial notification is sent.
- Quantity in commerce
- 30,366 cases of 20 units per case were distributed worldwide
- Distribution
- The product has been distributed to 454 consignees throughout the United States, Guam and Japan.
- Product codes and lots
- Model number 85863 (size 6 French), all lot codes beginning in 05 and 06