Lp(a) Calibrator, Cat No. 998-41491, 1 ml glass bottles, in cardboard box, 2 bottles/box ,Waco Chemicals USA...
Wako Chemicals, USA Inc began a recall of Lp(a) Calibrator, Cat No. 998-41491, 1 ml glass bottles, in cardboard box, 2 bottles/box ,Waco Chemicals USA, Inc., 1600 Bellwood Road Richmond, VA 23237... on . The reason FDA records is "Unexpected low cholesterol results-The Cholesterol calibrator (Lp(a)) component of the in vitro diagnostic kit for cholesterol testing, has provided patient results lower than the correct value". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1409-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- Lp(a) Calibrator, Cat No. 998-41491, 1 ml glass bottles, in cardboard box, 2 bottles/box ,Waco Chemicals USA, Inc., 1600 Bellwood Road Richmond, VA 23237, Manufactured by: Wako Pure Chemical Industries, Ltd. Osaka Japan
- Product code
- MMY Lipoprotein, Low Density, Removal
- Reason for recall
- Unexpected low cholesterol results-The Cholesterol calibrator (Lp(a)) component of the in vitro diagnostic kit for cholesterol testing, has provided patient results lower than the correct value.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The recalling firm notified consignees by phone and e-mail with an attached letter dated 07/06/06. The notification, flagged as ''Important Device Recall'' advised consignees to stop using the recalled product and that Lp(a) results showed a decrease in value of approximately 10% when the identified lots are used. Consignees were requested to return the enclosed response form for replacement product.
- Quantity in commerce
- 136 kits
- Distribution
- The product was sold to one distributor and direct medical accounts Nationwide.
- Product codes and lots
- Lots Numbers-EK560 EXP 7/31/06 and EH213 EXP 9/30/06