Varis Vision Treatment 6.6/RTP Exchange/Siemens Accelerators
Varian Medical Systems Oncology Systems began a recall of Varis Vision Treatment 6.6/RTP Exchange/Siemens Accelerators on . The reason FDA records is "A software anomaly may occur which can lead to patient treatment with the wrong field. The anomaly is reported only to occur when this software version (6.6.5022) is used with Elekta or Siemens linear accelerators". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1375-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Varis Vision Treatment 6.6/RTP Exchange/Siemens Accelerators
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- A software anomaly may occur which can lead to patient treatment with the wrong field. The anomaly is reported only to occur when this software version (6.6.5022) is used with Elekta or Siemens linear accelerators .
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 09/20/2005, Varian Medical Systems issued a notification letter via certified mail and FedEx. These letters were issued with instructions on what to do regarding the corrections. Varian Medical Systems intends to release new, updated software.
- Quantity in commerce
- 58 units
- Distribution
- The product was distributed to 34 consignees, located in AZ, CA, MA, MI, NC, PA in the USA, and in Australia, Belgium, Canada, Denmark, Finland, France, Germany, Italy, Lithuania, Spain, Sweden, and the United Kingdom
- Product codes and lots
- H462074, H462000, H462003, H462004, H462005, H462006, H462011, H462012, H462013, H462014, H462015, H462016, H462017, H462018, H462019, H462020, H462021, H462022, H462023, H462024, H462025, H462026, H462027, H462028, H462029, H462031, H462040, H462045, H462047, H462048, H462049, H462050, H462051, H462052, H462054, H462055, H462057, H462059, H462060, H462061, H462064, H462065, H462066, H462067, H462068, H462072, H462073, H462075, H462076, H462077, H462079, H462082, H462083, H462084, H462085, H462086, H462087