Z-1410-06

INOvent, Nitric Oxide Delivery System. Model Number 1605-9000-000

Mallinckrodt Manufacturing LLC / initiated 2006-06-29 / Terminated

INO Therapeutics, Inc. began a recall of INOvent, Nitric Oxide Delivery System. Model Number 1605-9000-000 on . The reason FDA records is "The Kel-F Tip mounted on the high pressure hose can become dislodge/lodged in the INOmax valve outlet". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1410-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Mallinckrodt Manufacturing LLC
Product
INOvent, Nitric Oxide Delivery System. Model Number 1605-9000-000
Product code
MRN Apparatus, Nitric Oxide Delivery
Reason for recall
The Kel-F Tip mounted on the high pressure hose can become dislodge/lodged in the INOmax valve outlet.
FDA root cause
Other
Action
On June 29, 2006, the firm sent a letter to its consignees advising them to visually verify that the Kel F tip is in place before attempting to connect either of the high pressure hoses to the INOmax cylinders. INO advised that if the tip was not in place, the operator should immediately replace the tip before attaching the hoses. On July 5, 2005, the firm sent a second letter to their consignees with extra Kel F tips and existing instructions from the INOvent Operation & Maintenance Manual outlining the proper replacement procedure for the tips. The firm is following with telephone contact to assure that the letters were delivered to the proper person(s).
Quantity in commerce
3,123 units
Distribution
Worldwide, USA and countries of Canada, Uruguay, Chile, Argentina, Colombia, and Malaysia
Product codes and lots
All serial numbers for Model #1605-9000-000