ADVIA 2120 systems -Automated Complete Blood Cell and Differential Cell Counter

Siemens Healthcare Diagnostics Inc. / initiated 2006-03-16 / Terminated

Bayer Healthcare, LLC (Diagnostics Division) began a recall of ADVIA 2120 systems -Automated Complete Blood Cell and Differential Cell Counter on . The reason FDA records is "The ADVIA 2120 has reported ; highly intermittent low results on all primary results: White Blood Cells (WBC), Red Blood Cells (RBC), Hemoglobin (HGB) and Platelets (PLT) and an abnormal baso cytogram". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1376-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA 2120 systems -Automated Complete Blood Cell and Differential Cell Counter
Product code
GKL Counter, Cell, Automated (Particle Counter)
Reason for recall
The ADVIA 2120 has reported ; highly intermittent low results on all primary results: White Blood Cells (WBC), Red Blood Cells (RBC), Hemoglobin (HGB) and Platelets (PLT) and an abnormal baso cytogram.
FDA root cause
Other
Action
Bayer HealthCare LLC issued Support Bulletins to all affected Bayer HealthCare LLC Branches on 3/16/06 in order to inform them of the issue and to provide them with instructions for this ''field correction''. Bayer also issued Customer Bulletins to the Branches both in the United States and elsewhere for communication with the affected customers.
Quantity in commerce
60 units Domestic, 258 units Foreign
Distribution
Worldwide distribution-Domestic consignees consist of hospitals and clinical laboratories. Foreign countries (units): Canada (1), China (4), Singapore (1), Malaysia (2), Thailand (1), Taiwan (1), Vietnam (1), Australia (3), Korea (25), Japan (25), Israel (4), Saudi Arabia (4), Russia (1), Turkey (1), United Arab Emirates (1), South Africa (2), United Kingdom (31), Germany (1), Poland (9), Norway (1), Spain (48), Finland (7), Denmark (4), Portugal (14), Italy (5), Netherlands (3), France (46), Ireland (4), India (7), Austria (1).
Product codes and lots
ADVIA 2120 Part No. 067-A012-01 (without Autosampler) and Part No. 067-A011-02 (with Autosampler); all ADVIA 2120 systems using Version 5 system software including Version 5.0.6 (Part No. 067-S045-06, English version), Version 5.1.1 (Part No. 067-S058-01, English version), and Version 5.2.7 (Part No. 067-S045-15, English version).