RapidLab 1200 Systems, Model 1245 - Blood gases, electrolyte and blood pH test system, Bayer HealthCare LLC...
Siemens Healthcare Diagnostics began a recall of RapidLab 1200 Systems, Model 1245 - Blood gases, electrolyte and blood pH test system, Bayer HealthCare LLC, Diagnostics Division on . The reason FDA records is "Invalid CO-oximeter values; Bayer HealthCare determined that under very specific conditions, the RapidLab 1245 or RapidLab 1265 systems may report invalid CO-oximeter values to an LIS, Rapidlink, or Rapidcomm™ data...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1464-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- RapidLab 1200 Systems, Model 1245 - Blood gases, electrolyte and blood pH test system, Bayer HealthCare LLC, Diagnostics Division
- Product code
- CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
- Reason for recall
- Invalid CO-oximeter values; Bayer HealthCare determined that under very specific conditions, the RapidLab 1245 or RapidLab 1265 systems may report invalid CO-oximeter values to an LIS, Rapidlink, or Rapidcomm™ data management system.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 4/04/06, Bayer HealthCare LLC (Diagnostics Division) sent Support Bulletins to all affected Bayer HealthCare LLC Branches via e-mail in order to inform them of the issue, and to provide them with instructions for the interim work-around until the new software version is installed. Customer Bulletins were prepared and sent to the Branches both in the United States and globally for communication with the affected customers.
- Quantity in commerce
- Domestic - 124 units, Foreign-364 units (all systems)
- Distribution
- Worldwide
- Product codes and lots
- Part No. 05061537; all units using Version 1.0, 1.01, or 1.02 software connected to LIS, Rapidlink¿, or Rapidcomm™ data management systems.