Z-1401-06
LANTIS XLINK (Impac MultiAccess Sequencer) on Varian accelerators with the MLC3PI interface
Siemens Medical Solutions USA, Inc. / initiated 2006-07-07 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of LANTIS XLINK (Impac MultiAccess Sequencer) on Varian accelerators with the MLC3PI interface on . The reason FDA records is "Improper treatment may be delivered if the machine configuration file is edited/set for the Varian MLC revision H file format". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1401-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- LANTIS XLINK (Impac MultiAccess Sequencer) on Varian accelerators with the MLC3PI interface
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Improper treatment may be delivered if the machine configuration file is edited/set for the Varian MLC revision H file format.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm has issued a safety advisory letter to all affected consignees, and will distribute an updated software version as soon as it is available.
- Quantity in commerce
- 23 units
- Distribution
- Worldwide Distribution --- USA and countries including Australia, China, Denmark, Germany, New Zealand, Portugal, Russia, Singapore, South Africa, Sweden and Switzerland
- Product codes and lots
- Site Numbers: 93-0452, 93-0494, 93-0571, 93-0724, 93-0881, 93-0894, 93-0902, 93-0443, 93-0776, 93-0677, 93-1032, 93-0880, 93-0010, 93-0095, 93-0432, 93-0991, 93-0711, 93-0007, 93-0641, 93-0356, 93-0391, 93-0015, 93-0118