Z-1506-06

Ruggles Leyla Ball Joint Clamp, MODEL/CATALOG #: R2383 . The Leyla Ball Joint Clamp is part of the Table...

J. Jamner Surgical Instruments, Inc. / initiated 2006-04-24 / Terminated

J. Jamner Surgical Instruments, Inc. began a recall of Ruggles Leyla Ball Joint Clamp, MODEL/CATALOG #: R2383 . The Leyla Ball Joint Clamp is part of the Table Mounting Hardware that is used with the Leyla... on . The reason FDA records is "The Ruggles Leyla Ball Joint Clamp may not properly tighten onto the Rigid Holding Rod. Although this condition can be observed during assembly, if the device were to be used during surgery, any attachments to the Rigid...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1506-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
J. Jamner Surgical Instruments, Inc.
Product
Ruggles Leyla Ball Joint Clamp, MODEL/CATALOG #: R2383 . The Leyla Ball Joint Clamp is part of the Table Mounting Hardware that is used with the Leyla Retractor System. The Leyla Table Mounting Hardware consists of the Ball Joint Clamp, the Rigid Holding Rod, the Coupling Head, and the Coupling Head Turn Table. The Ball Joint Clamp is available individually or as part of the Table Mount Set. The Table Mounting Hardware are attachments to operating room surgical tables. The product is supplied non-sterile and may be packaged individually or as part of the Table Mount Set.
Product code
FWZ Operating Room Accessories Table Tray
Reason for recall
The Ruggles Leyla Ball Joint Clamp may not properly tighten onto the Rigid Holding Rod. Although this condition can be observed during assembly, if the device were to be used during surgery, any attachments to the Rigid Holding Rod could move.
FDA root cause
Other
Action
On 4/24/06, Integra LifeSciences Corp. sent an 'URGENT - RECALL NOTICE' letter (dated 4/24/06) and a 'RECALL ACKNOWLEDGEMENT' Form to the 5 hospitals and 1 salesman that received the product.
Quantity in commerce
8 units
Distribution
5 hospitals in MS, IL, WI and TX, and 1 salesman in IL.
Product codes and lots
The affected product can be identified by the following code on the product label: 143 60173 6.