Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1468-06

Zimmer VERSYS Hip System Femoral Stem, press-fit, enhanced taper, 12/14 neck taper, collarless, size 14, 135...

Zimmer GmbH / initiated 2006-08-02 / Terminated

Zimmer Inc. began a recall of Zimmer VERSYS Hip System Femoral Stem, press-fit, enhanced taper, 12/14 neck taper, collarless, size 14, 135 mm stem length, tivanium TI-6AL-4V alloy, sterile... on . The reason FDA records is "Lack of assurance of proper metal fatigue strength due to a metal grain structure anomaly". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1468-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer VERSYS Hip System Femoral Stem, press-fit, enhanced taper, 12/14 neck taper, collarless, size 14, 135 mm stem length, tivanium TI-6AL-4V alloy, sterile; Cat. no. 7861-14-04 (00786101404).
Product code
LWJ Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented
Reason for recall
Lack of assurance of proper metal fatigue strength due to a metal grain structure anomaly.
FDA root cause
Other
Action
Consignees were notified via letter dated 8/2/06 to cease use of and to return the products. Zimmer distributors were to examine stock and return product on hand.
Quantity in commerce
181 of all devices
Distribution
Worldwide, including USA, Australia, China, Denmark, Japan, Jordan, Latin Ameria, Russia, Singapore and Switzerland.
Product codes and lots
Lots 60454196, 60454197, 60454198, 60454199, 60465016 and 60465017.