Alaris SE Pump (formerly Signature Edition Infusion Pump); Model Numbers: 7000, 7100, 7101, 7130, 7131, 7132...
Cardinal Health 303 Inc DBA Alaris Products began a recall of Alaris SE Pump (formerly Signature Edition Infusion Pump); Model Numbers: 7000, 7100, 7101, 7130, 7131, 7132, 7200, 7201, 7230, 7231, and 7232 on . The reason FDA records is "This recall was initiated because of a potential for over infusion with all models of the Alaris SE Pumps (formerly the Signature Edition Infusion Pumps) caused by key bounce". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1484-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- Alaris SE Pump (formerly Signature Edition Infusion Pump); Model Numbers: 7000, 7100, 7101, 7130, 7131, 7132, 7200, 7201, 7230, 7231, and 7232
- Product code
- FRN Pump, Infusion
- Reason for recall
- This recall was initiated because of a potential for over infusion with all models of the Alaris SE Pumps (formerly the Signature Edition Infusion Pumps) caused by key bounce.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Corrective action for this recall will include notifications (five letters) being sent (Return Receipt) to all customers with affected product alerting them to this action. The letters are addressed to the Director of Biomedical Engineering, Director, Materials Management, Hospital Administration, Purchasing, and the fifth letter to the Director of Nursing. The August 15, 2006, recall letters and the warning labels will be sent return receipt requested. The verification of receipt of the notifications will be tracked and return receipt exceptions will be followed up. Follow up activities will be documented including customer contact.
- Quantity in commerce
- 147, 461
- Distribution
- Nationwide and American Samoa, Guam, and Puerto Rico. Worldwide.
- Product codes and lots
- Model Numbers: 7000, 7100, 7101, 7130, 7131, 7132, 7200, 7201, 7230, 7231, and 7232. (ALL MODELS, ALL LOTS) The serial numbers are not sequential.