Z-1535-06
NICO Model 7300 Cardiopulmonary Management System
Respironics Novametrix, LLC. / initiated 2006-08-11 / Terminated
Respironics Novametrix, LLC. began a recall of NICO Model 7300 Cardiopulmonary Management System on . The reason FDA records is "Audible alarm may not sound during an alert condition". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1535-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Respironics Novametrix, LLC.
- Product
- NICO Model 7300 Cardiopulmonary Management System
- Product code
- CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
- Reason for recall
- Audible alarm may not sound during an alert condition
- FDA root cause
- Other
- Action
- Respironics notified consignees of the problem with the audible alarm by certified letter on 8/28/06. The correction requires replacement of a speaker upgrade kit.
- Quantity in commerce
- 2,432 units
- Distribution
- Worldwide distribution --- USA and countries of United Arab Emirates, Argentina, Austria, Australia, Belgium, Bulgaria, Bahrain, Brazil, Canada, Canton of Aargau Switzerland, Chile, China, Columbia, Costa Rica, Czech Republic, Germany, Denmark, Algeria, Ecuador, Egypt, Spain, Finland, France, Great Britain, Guam, Hong Kong, Hungary, Ireland, India, Italy, Jamaica, Jordon, Japan, Korea, Kuwait, Lebanon, Lithuania, Morocco, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Peru, Phillippines, Puerto Rico, Portugal, Qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovania, Slovakia, Thailand, Turkey, Taiwan, Uruguay, Venezuala, and South Africa.
- Product codes and lots
- All serial numbers beginning with 138