Z-1535-06

NICO Model 7300 Cardiopulmonary Management System

Respironics Novametrix, LLC. / initiated 2006-08-11 / Terminated

Respironics Novametrix, LLC. began a recall of NICO Model 7300 Cardiopulmonary Management System on . The reason FDA records is "Audible alarm may not sound during an alert condition". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1535-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Respironics Novametrix, LLC.
Product
NICO Model 7300 Cardiopulmonary Management System
Product code
CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Reason for recall
Audible alarm may not sound during an alert condition
FDA root cause
Other
Action
Respironics notified consignees of the problem with the audible alarm by certified letter on 8/28/06. The correction requires replacement of a speaker upgrade kit.
Quantity in commerce
2,432 units
Distribution
Worldwide distribution --- USA and countries of United Arab Emirates, Argentina, Austria, Australia, Belgium, Bulgaria, Bahrain, Brazil, Canada, Canton of Aargau Switzerland, Chile, China, Columbia, Costa Rica, Czech Republic, Germany, Denmark, Algeria, Ecuador, Egypt, Spain, Finland, France, Great Britain, Guam, Hong Kong, Hungary, Ireland, India, Italy, Jamaica, Jordon, Japan, Korea, Kuwait, Lebanon, Lithuania, Morocco, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Peru, Phillippines, Puerto Rico, Portugal, Qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovania, Slovakia, Thailand, Turkey, Taiwan, Uruguay, Venezuala, and South Africa.
Product codes and lots
All serial numbers beginning with 138