Z-0305-2007

Histochemical PAS (Periodic Acid Schiff) Reaction Set. in vitro diagnostic. Product size configuration: Set...

EMD Millipore Corporation / initiated 2006-07-25 / Terminated

EMD Chemicals Inc began a recall of Histochemical PAS (Periodic Acid Schiff) Reaction Set. in vitro diagnostic. Product size configuration: Set of 4 bottles including Schiff Reagent 225mL, Light... on . The reason FDA records is "Due to loose caps, the SO2 levels dropped significantly from release levels causing the product not to perform properly". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0305-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
EMD Millipore Corporation
Product
Histochemical PAS (Periodic Acid Schiff) Reaction Set. in vitro diagnostic. Product size configuration: Set of 4 bottles including Schiff Reagent 225mL, Light Green SF Yellowish Stain 225mL, Sodium Carbonate Solution 225mL, and Periodic Acid Solution 225mL. Each kit provides reagents sufficient for a maximum of 100 tests.
Product code
KKS Acid, Periodic
Reason for recall
Due to loose caps, the SO2 levels dropped significantly from release levels causing the product not to perform properly.
FDA root cause
Other
Action
On July 27, 2006 EMD contacted its direct account customers via letter. The firm requested discontinue use of the product and contact EMD for return.
Quantity in commerce
74 kits
Distribution
Nationwide.
Product codes and lots
Item Number 64945/93 Lot # 5341, Exp. Date: May 31, 2007.