Z-1503-06
Hill-Rom Affinity Three Birthing Bed, Model P3700
Hill-Rom Inc. / initiated 2006-08-18 / Terminated
Hill-Rom, Inc. began a recall of Hill-Rom Affinity Three Birthing Bed, Model P3700 on . The reason FDA records is "The cable on the auxiliary outlet may become pinched, which may result in an electrical short with melting of the plastic transformer and the emission of smoke". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1503-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- Hill-Rom Affinity Three Birthing Bed, Model P3700.
- Product code
- KNC Table, Obstetric (And Accessories)
- Reason for recall
- The cable on the auxiliary outlet may become pinched, which may result in an electrical short with melting of the plastic transformer and the emission of smoke.
- FDA root cause
- Other
- Action
- Customers were notified via letter dated 8/18/06 and instructed to either correct the problem themselves with the instructions provided or to contact the firm requesting a visit from the recalling firm to correct their beds.
- Quantity in commerce
- 9,324
- Distribution
- Nationwide, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, France, French Polynesia, Germany, Guam, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Republic of Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Sait Kitts and Nevis, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, U.S. Virgin Islands and Venezuela.
- Product codes and lots
- All units distributed after October 2000 and equipped with the 12VAC accessory outlet option.