Z-1513-06
Stryker Power-PRO Powered Ambulance Cot, Model 6500
Stryker Medical / initiated 2006-08-29 / Terminated
Stryker Medical Div. of Stryker Corporation began a recall of Stryker Power-PRO Powered Ambulance Cot, Model 6500 on . The reason FDA records is "The head section wheel assembly may flex outward from the cot because the head section wheel bolts, which hold the assembly upright, may back out". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1513-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Medical
- Product
- Stryker Power-PRO Powered Ambulance Cot, Model 6500.
- Product code
- FPO Stretcher, Wheeled
- Reason for recall
- The head section wheel assembly may flex outward from the cot because the head section wheel bolts, which hold the assembly upright, may back out.
- FDA root cause
- Other
- Action
- Consignees were notified of the recall via letter dated 8/29/06. Included with the letter were replacement bolts for each cot and installation instructions, but the customer was given the option of contacting the recalling firm for installation.
- Quantity in commerce
- 1,841 cots (44 foreign)
- Distribution
- Worldwide distribution ---- including USA and countries of Canada, China, Germany and Switzerland.
- Product codes and lots
- All units with a serial number lower than 060739380. (units manufactured between 1 Oct 2005 and 15 July 2006)