Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1513-06

Stryker Power-PRO Powered Ambulance Cot, Model 6500

Stryker Medical / initiated 2006-08-29 / Terminated

Stryker Medical Div. of Stryker Corporation began a recall of Stryker Power-PRO Powered Ambulance Cot, Model 6500 on . The reason FDA records is "The head section wheel assembly may flex outward from the cot because the head section wheel bolts, which hold the assembly upright, may back out". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1513-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Medical
Product
Stryker Power-PRO Powered Ambulance Cot, Model 6500.
Product code
FPO Stretcher, Wheeled
Reason for recall
The head section wheel assembly may flex outward from the cot because the head section wheel bolts, which hold the assembly upright, may back out.
FDA root cause
Other
Action
Consignees were notified of the recall via letter dated 8/29/06. Included with the letter were replacement bolts for each cot and installation instructions, but the customer was given the option of contacting the recalling firm for installation.
Quantity in commerce
1,841 cots (44 foreign)
Distribution
Worldwide distribution ---- including USA and countries of Canada, China, Germany and Switzerland.
Product codes and lots
All units with a serial number lower than 060739380. (units manufactured between 1 Oct 2005 and 15 July 2006)