Z-0412-2007
Thoratec Dual Drive Console a drive component of the Thoratec Ventricular Assist Device, Model Numbers...
Abbott Medical / initiated 2004-09-24 / Terminated
Thoratec Corp began a recall of Thoratec Dual Drive Console a drive component of the Thoratec Ventricular Assist Device, Model Numbers, 10025-2600-005, 10025-2601-007, 10025-2602-006... on . The reason FDA records is "The module power switch (at the back of console) may break while attempting to turn the Dual Drive Console OFF or ON". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0412-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Thoratec Dual Drive Console a drive component of the Thoratec Ventricular Assist Device, Model Numbers, 10025-2600-005, 10025-2601-007, 10025-2602-006, Thoratec Corporation, 6035 Stoneridge Drive, Pleasanton, CA 94588
- Product code
- DSQ Ventricular (Assist) Bypass
- Reason for recall
- The module power switch (at the back of console) may break while attempting to turn the Dual Drive Console OFF or ON.
- FDA root cause
- Other
- Action
- The firm issued on 1/25/07 an URGENT: MEDICAL DEVICE CORRECTION-notice informing its consignees about the 'Power Switch Replacement Instructions' and arrangements for their service department replacement of the switches.
- Quantity in commerce
- 331
- Distribution
- Worldwide
- Product codes and lots
- All Dual Drive Consoles manufactured before July 10, 2002 with serial numbers between 301 and 636, inclusive