Z-0412-2007

Thoratec Dual Drive Console a drive component of the Thoratec Ventricular Assist Device, Model Numbers...

Abbott Medical / initiated 2004-09-24 / Terminated

Thoratec Corp began a recall of Thoratec Dual Drive Console a drive component of the Thoratec Ventricular Assist Device, Model Numbers, 10025-2600-005, 10025-2601-007, 10025-2602-006... on . The reason FDA records is "The module power switch (at the back of console) may break while attempting to turn the Dual Drive Console OFF or ON". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0412-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Thoratec Dual Drive Console a drive component of the Thoratec Ventricular Assist Device, Model Numbers, 10025-2600-005, 10025-2601-007, 10025-2602-006, Thoratec Corporation, 6035 Stoneridge Drive, Pleasanton, CA 94588
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
The module power switch (at the back of console) may break while attempting to turn the Dual Drive Console OFF or ON.
FDA root cause
Other
Action
The firm issued on 1/25/07 an URGENT: MEDICAL DEVICE CORRECTION-notice informing its consignees about the 'Power Switch Replacement Instructions' and arrangements for their service department replacement of the switches.
Quantity in commerce
331
Distribution
Worldwide
Product codes and lots
All Dual Drive Consoles manufactured before July 10, 2002 with serial numbers between 301 and 636, inclusive