Z-1550-06

Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03253678122

Roche Diagnostics GmbH / initiated 2006-09-15 / Terminated

Roche Diagnostics Corp. began a recall of Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03253678122 on . The reason FDA records is "Gel-like particle formation; may result in a clot settling in the analyzer measuring cell, which will create a low signal and may cause incorrectly high or low QC and/or patient results depending on the assay method". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1550-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03253678122.
Product code
CGN Acid, Folic, Radioimmunoassay
Reason for recall
Gel-like particle formation; may result in a clot settling in the analyzer measuring cell, which will create a low signal and may cause incorrectly high or low QC and/or patient results depending on the assay method.
FDA root cause
Other
Action
Consignees were informed via letter dated 9/15/06 to perform Folate assays in batch mode and to run controls and liquid flow cleaning upon completion of each batch run. Distributors were requested to notify their customers.
Quantity in commerce
5688 units
Distribution
Nationwide.
Product codes and lots
Lots 175646 and 174558.