GE Precision 500D Radiographic and Fluoroscopic Imaging System, Stationary X-ray System consisting of an...
General Electric Med Systems LLC began a recall of GE Precision 500D Radiographic and Fluoroscopic Imaging System, Stationary X-ray System consisting of an X-ray generator; angulating table with x-ray Tube... on . The reason FDA records is "Systems are non-compliant with Federal Performance Standard Title 21, Code of Federal Regulation (CFR), 1020.32(a)(1) in that the systems allow the production of x-rays when the primary protective barrier is not in...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1304-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Precision 500D Radiographic and Fluoroscopic Imaging System, Stationary X-ray System consisting of an X-ray generator; angulating table with x-ray Tube, collimator and image intensifier; wall stand; Overhead tube suspension; operator console and digital archive system.
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- Systems are non-compliant with Federal Performance Standard Title 21, Code of Federal Regulation (CFR), 1020.32(a)(1) in that the systems allow the production of x-rays when the primary protective barrier is not in position to intercept the entire cross section of the useful beam
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- An Important Notice for Precision 500D, dated 11/24/2005 was hand delivered to all customers beginning 05/15/06, by GE field service engineers at the time they were to perform the correction to the device.
- Quantity in commerce
- 1,289
- Distribution
- Worldwide, including USA, Puerto Rico, Canada, France, Germany, Indonesia, Italy, Japan, Jordan, Kuwait, Macedonia, Panama, Russia, Saudi Arabia, Singapore, and United Kingdom.
- Product codes and lots
- All serial numbers affected