Z-1304-06

GE Precision 500D Radiographic and Fluoroscopic Imaging System, Stationary X-ray System consisting of an...

GE Healthcare / initiated 2006-05-15 / Terminated / root cause: manufacturing process

General Electric Med Systems LLC began a recall of GE Precision 500D Radiographic and Fluoroscopic Imaging System, Stationary X-ray System consisting of an X-ray generator; angulating table with x-ray Tube... on . The reason FDA records is "Systems are non-compliant with Federal Performance Standard Title 21, Code of Federal Regulation (CFR), 1020.32(a)(1) in that the systems allow the production of x-rays when the primary protective barrier is not in...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1304-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Precision 500D Radiographic and Fluoroscopic Imaging System, Stationary X-ray System consisting of an X-ray generator; angulating table with x-ray Tube, collimator and image intensifier; wall stand; Overhead tube suspension; operator console and digital archive system.
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
Systems are non-compliant with Federal Performance Standard Title 21, Code of Federal Regulation (CFR), 1020.32(a)(1) in that the systems allow the production of x-rays when the primary protective barrier is not in position to intercept the entire cross section of the useful beam
Root cause tag
manufacturing process
FDA root cause
Other
Action
An Important Notice for Precision 500D, dated 11/24/2005 was hand delivered to all customers beginning 05/15/06, by GE field service engineers at the time they were to perform the correction to the device.
Quantity in commerce
1,289
Distribution
Worldwide, including USA, Puerto Rico, Canada, France, Germany, Indonesia, Italy, Japan, Jordan, Kuwait, Macedonia, Panama, Russia, Saudi Arabia, Singapore, and United Kingdom.
Product codes and lots
All serial numbers affected