Z-0004-2007
Thoratec brand TLC-II Portable Ventricular Assist Device (VAD) Driver; Catalog No. 20010-2085-000 Distributed...
Abbott Medical / initiated 2006-08-25 / Terminated
Thoratec Corp began a recall of Thoratec brand TLC-II Portable Ventricular Assist Device (VAD) Driver; Catalog No. 20010-2085-000 Distributed and Manufactured by Thoratec Corporation, 6035... on . The reason FDA records is "Sticky Valve Disk- The valve disk has an increased tendency to stick during operation causing noise and pressure/vacuum alarms". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0004-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Thoratec brand TLC-II Portable Ventricular Assist Device (VAD) Driver; Catalog No. 20010-2085-000 Distributed and Manufactured by Thoratec Corporation, 6035 Stoneridge Drive, Pleasanton, CA 94588
- Product code
- DSQ Ventricular (Assist) Bypass
- Reason for recall
- Sticky Valve Disk- The valve disk has an increased tendency to stick during operation causing noise and pressure/vacuum alarms.
- FDA root cause
- Other
- Action
- On 8/26/06, the firm issued letters to all its direct consignees, informing them of the affected product and providing 'IMPORTANT PRODUCT SAFETY BULLETIN'.
- Quantity in commerce
- 114 devices (57 valves/2 per unit)
- Distribution
- Worldwide distribution-USA including states of AL, CA, CT, FL, IL, IN, KY, LA, MA, ME, MI, MN, NC, NY, OH, PA, TN, TX, VA, WA, and WI., with 19 Foreign consignees Australia, Belgium, Canada*, Czech Republic, France, Germany, Ireland, Netherlands, Portugal, Switzerland, United Kingdom. (*Hospital Royal Victoria, Queen Elizabeth II/Capital Health, St. Paul's Hospital, Vancouver, Canada.)--
- Product codes and lots
- All Codes (Valve Disk Lot 20052904)