Z-0201-2007
Datex-Ohmeda Tec 6 Plus Desflurane Vaporizer. It is an electronic vaporizer which delivers the anesthetic...
GE Healthcare / initiated 2006-09-15 / Terminated
Datex - Ohmeda, Inc began a recall of Datex-Ohmeda Tec 6 Plus Desflurane Vaporizer. It is an electronic vaporizer which delivers the anesthetic agent desflurane on . The reason FDA records is "There is a possibility of a vaporizer power failure, with audible and visual alarms, when using the Tec 6 Plus Desflurane Vaporizer". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0201-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Datex-Ohmeda Tec 6 Plus Desflurane Vaporizer. It is an electronic vaporizer which delivers the anesthetic agent desflurane.
- Product code
- CAD Vaporizer, Anesthesia, Non-Heated
- Reason for recall
- There is a possibility of a vaporizer power failure, with audible and visual alarms, when using the Tec 6 Plus Desflurane Vaporizer.
- FDA root cause
- Other
- Action
- An Urgent Medical Device Correction letter dated September 15, 2006, was sent to affected customers. The letter describes the issue, states that a GE representative will contact affected accounts to arrange for the vaporizer exchange at no charge.
- Quantity in commerce
- 93
- Distribution
- Nationwide and Canada.
- Product codes and lots
- serial numbers: AETK25021 thru AETK25023, AETK25024, AETK25025, AETK25027 thru AETK25030, AETK27001, AETK27002, AETK27004, AETK27006, AETK27007, AETK27010 thru AETK27015, AETK27017, AETK27019, AETK27020, AETK27024, AETK27025, AETK27027, AETK27031 thru AETK27034, AETK27039, AETK27040, AETK28001 thru AETK28003, AETK28005, AETK28006, AETK28008, AETK28010, AETK28011, AETK28013 thru AETK28017, AETK28019, AETK28021, AETK28023 thru AETK28025, AETK28029, AETK28031thru AETK28038, AETK28040 thru AETK28046, AETK28048, AETK28050 thru AETK28052, AETK28056, AETK28057, AETK28059, AETK28063, AETK28065, AETK28067, AETK28070 thru AETK28072, AETK28074 thru AETK28076, AETK29004, AETK29006, AETK29035, AETK29037, AETK29039, AETK29043, AETK29045, AETK29061, AETK29068, AETK29073, AEWK30001.