Z-1517-06
Terumo Khuri Myocardial Ph Monitoring System, Catalog Number 7205
Terumo Cardiovascular Systems Corporation / initiated 2006-06-23 / Terminated / root cause: sterility
Terumo Cardiovascular Systems Corp began a recall of TERUMO KHURI MYOCARDIAL PH MONITORING SYSTEM, CATALOG NUMBER 7205 on . The reason FDA records is "Lack of Assurance of Sterility: these specific lots of sensors lack supporting documentation for sterilization assurance". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1517-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Khuri Myocardial Ph Monitoring System, Catalog Number 7205
- Product code
- CBZ Analyzer, Ion, Hydrogen-Ion (Ph), Blood-Phase, Indwelling
- Reason for recall
- Lack of Assurance of Sterility: these specific lots of sensors lack supporting documentation for sterilization assurance.
- Root cause tag
- sterility
- FDA root cause
- Other
- Action
- The affected lots were identified as those lots manufactured in the first quarter of the calendar year 2006. The distribution of affected lots was traced. A letter and a customer response form were developed and each of the five US customers were sent the recall letter via Federal Express. The two international affiliates were contacted and instructed to contact their customers in a manner similar to one used by Terumo CVS in the US. Reconciliation will involve confirmation that the response form for each of the customers and affiliates has been completed and returned to Terumo CVS.
- Quantity in commerce
- 175
- Distribution
- Nationwide in CA, IL, MA, MD, NC, & OK; Internationally in Belgium & Kuwait
- Product codes and lots
- Lots Affected: 2006-O6AD, 2006-O8AD, 2006-O8AP, 6O8AP6O8AD