Z-0158-2007
Jostra - HL-20 Heart Lung Integrated Perfusion System, Roller Pump Hand crank
Maquet Cardiovascular Us Sales, LLC / initiated 2006-09-01 / Terminated
MAQUET Inc. began a recall of Jostra - HL-20 Heart Lung Integrated Perfusion System, Roller Pump Hand crank on . The reason FDA records is "The hand crank necessary for the emergency manual operation of the pump is too wide diametrically to fit into the holes which they were designed to fill". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0158-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Jostra - HL-20 Heart Lung Integrated Perfusion System, Roller Pump Hand crank
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- The hand crank necessary for the emergency manual operation of the pump is too wide diametrically to fit into the holes which they were designed to fill.
- FDA root cause
- Other
- Action
- The firm sent a safety alert letter on 9/13/06 to all US customers by Federal Express Return Receipt.
- Quantity in commerce
- 132 or less in US. Firm officials stated that 66 base units in US 8/1/05 to 8/14/06 - each units comes with 2 hand cranks.
- Distribution
- worldwide
- Product codes and lots
- Article # 923391 The affected HL-20 base units include S/Ns 100001 through 100240. Hand cranks shipped between the dates of 8/15/05 to 8/14/06 have the potential to be defective. Serial #s100155, 1025, 1026, 1032, 1331, 1332, 100011, 100015, 100016, 100017, 1272, 1417, 1480, 1500, 1653, 14031054, 14031055, 14031059, 14041074, 14041874, 100012, 100013, J14041874, 1584, 100108, 100109, 14041074, JOSTRA, 100009, 100010, 100130, 100131, 1625, 1626, 1627, 1628, 1629, 100135, 100136, 100137, 14031040, 14031068, 100188, 1019, 1020, 1021, 1295, 100021, 100022, 100116, 14031051, 14031052, 1507, 1508, 1509, 100017, 100219, 100220, 100221, 100222, 100223, 100224, 100225, 100226, 100227, 100228.