Z-0159-2007
Jostra HL-20 Heart Lung Machine - guiding pins
Maquet Cardiovascular Us Sales, LLC / initiated 2006-10-03 / Terminated
MAQUET Inc. began a recall of Jostra HL-20 Heart Lung Machine - guiding pins on . The reason FDA records is "The HL-20's pump head on the roller pump has four pairs of tube guide rolls that keep the tube in place in the raceway and roll along the inserted tube as the pump head turns. If dirt and other particles accumulate in...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0159-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Jostra HL-20 Heart Lung Machine - guiding pins
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- The HL-20's pump head on the roller pump has four pairs of tube guide rolls that keep the tube in place in the raceway and roll along the inserted tube as the pump head turns. If dirt and other particles accumulate in the tube guide rolls, they cannot roll smoothly and finally they can get stuck.
- FDA root cause
- Other
- Action
- An 'Important Safety Notice - Please Read Immediately' letter was sent to all US customers on 10/3/06 by FedEx Return Receipt.
- Quantity in commerce
- 227 HL-20 base units with 545 single pumps
- Distribution
- Worldwide
- Product codes and lots
- All HL 20 systems Serial #s 1159,14031794,14031795, 14031796,14031797,14031798, 100155, 1345, 1363,1588, 14031765, 14031766, 1570, 1571,1593,1319,1320,1321,1322,14031774,14031775, 14031776, 1323,1324,1478,1479,14031067,14031069,10331385,1597,14021730,14031792,1459,1701,1512,1513,100035, 100036, 1034,1035, 14031045, 14031046, 14031047, 14031048, 1022,1023,1541,1559,100094,1025,1026,1032,1331,1332, 100011,100015,100016,100009,100063, 1668,1511, 1652, 1702,1454,1455, 1379,1380, 1381, 1272, 1417,1480, 1500, 1653, 14031054, 14031055,14031059, 14041074, 14041874, 1128, 1592, 100030, 100031, 100012, 100013, 1595, 1596, 14031039, 1510, 1105, 1433, 1434, 1456, 1457, 1577, 1578, 1579, 14021718, 14021719, 14041904, 100026, 14021706, 1420, 1421, 1398, 1418, 1598, J14041874,1584, 100108, 100109,14041074, JOSTRA, 1132, 100009, 100010,100130,100131, 1625, 1626, 1627, 1628, 1629, 100135, 100136, 100137, 1199, 1200, 1540, 1552, 1567, 1568, 1569, 1580, 1581, 1584, 1585, 1586, 14031040, 14031068, 14031761, 14031762, 1716, 100084, 14021716, 14041882, 100188, 1562, 1563, 100028, 100023, 14031831, 14031793, 1019, 1020, 1021, 1295, 100021, 100022, 14031777, 14031778, 14031779, 1523, 1524, 1017, 1442, 1443, 1465, 1466, 1635, 100116, 14031051, 14031052, 1507, 1508, 1509, 14031732, 100001, 100002, 100061, 100062, 14031734, 14031735, 14031736, 100017, 1483, 1481, 1482, 14031056, 14031057, 1535,1536, 1537, 1538, 1539, 1547, 1548, 1549, 1550, 1551, 14031060, 14031061, 14031062, 14031063, 100219,100220, 100221, 100222, 100223, 100224, 100225, 100226, 100227, 100228, 1642, 1643, 1494, 1666, 1667, 1618, 1619, 1517, 1518, 100027, 1345