BD Vacutainer Push Button Blood Collection Set with Pre-attached Holder. Non-pyrogenic 100. Rx only, sterile...
Becton Dickinson & Company began a recall of BD Vacutainer Push Button Blood Collection Set with Pre-attached Holder. Non-pyrogenic 100. Rx only, sterile REF 368656 23G x 3/4'' x 12'' 0.6 x 19mm x 305mm... on . The reason FDA records is "A complaint was received regarding a BD blood collection set where the blister was noticeably warped. The seal width was below minimum specification. Some blisters could adhere together and crack upon separation. Some...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0034-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Vacutainer Push Button Blood Collection Set with Pre-attached Holder. Non-pyrogenic 100. Rx only, sterile REF 368656 23G x 3/4'' x 12'' 0.6 x 19mm x 305mm ----- REF 367352 21G x 3/4'' x 12'' 0.8 x 19mm x 305mm
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- A complaint was received regarding a BD blood collection set where the blister was noticeably warped. The seal width was below minimum specification. Some blisters could adhere together and crack upon separation. Some units may have pre-activated or have defects affecting package integrity.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Recall notification were mailed out first class mail, return receipt requested, approximately 8/24/2006 to distributors, who were then asked to contact their customers.
- Quantity in commerce
- 110,300 units
- Distribution
- Nationwide.
- Product codes and lots
- Catalog number 367352 Lot # 6100596 Catalog number 367352 Lot # 6100597 Catalog number 367352 Lot # 6152413 Catalog number 368656 Lot # 6090251 Catalog number 368656 Lot # 6094813 Catalog number 368656 Lot # 6094840 Catalog number 368656 Lot # 6111757 Catalog number 368656 Lot # 6121663 Catalog number 368656 Lot # 6146186