Medtronic Intersept Custom Tubing Pack. Do Not Reuse. The design and components of this product were...
Medtronic Perfusion Systems began a recall of Medtronic Intersept Custom Tubing Pack. Do Not Reuse. The design and components of this product were specified by the user from an array of components and... on . The reason FDA records is "An out of specification condition exists with the tapered tubing found in some Custom Tubing Packs. The wall thickness is undersized and does not meet specifications". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0340-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Perfusion Systems
- Product
- Medtronic Intersept Custom Tubing Pack. Do Not Reuse. The design and components of this product were specified by the user from an array of components and designs offered by the manufacturer. Any change in the cardiopulmonary bypass procedure that compromises the function of this Custom Tubing Pack is the responsibility of the user. The tubing is made of polyvinyl chloride (PVC). Portions of the tapered tubing are used in the vent, suction and arterial roller heads of the heart lung machine. Manufacturer: Medtronic, Inc., Minneapolis, MN 55432-5604 USA. Tube supplier: Natvar (Tekni-Plex Co.), 19555 E. Arenth Ave., City of Industry, CA 91748.
- Product code
- DWE Tubing, Pump, Cardiopulmonary Bypass
- Reason for recall
- An out of specification condition exists with the tapered tubing found in some Custom Tubing Packs. The wall thickness is undersized and does not meet specifications.
- FDA root cause
- Other
- Action
- An Important Customer Notification, dated 09/06/06, was sent to affected customers. The letter identifies the affected model and lot numbers. Medtronic requests that affected product be returned. A suplemental pack can be built and shipped to the customer upon request. Medtronic will replace all returned product. A Recall Certificate is asked to be completed by customer and faxed to Medtronic.
- Quantity in commerce
- 146
- Distribution
- Distributed in 6 states: CA, FL, IL, IN, LA and TN
- Product codes and lots
- The following Model / Lot Number are affected: CB4W39R1 / 4876667; 0A84R2 / 4531036; 1708R18 / 4627100; 0V68R4 / 4212182; 1P07R6 / 4433014; 1P07R6 / 4678134; 2Y91R / 4628170; 2Y91R / 4800224; 2Y91R / 4902036; CB0Q49R14 / 4800849; 4D87R2 / 4184927.