Z-0169-2007
Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal...
Baxter Healthcare Renal Div / initiated 2006-09-06 / Terminated
Baxter Healthcare Renal Div began a recall of Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A on . The reason FDA records is "The Meridian pump is less likely to detect small pressure changes, which may indicate the presence of a post-pump tubing kink, when using pre-pump arterial pressure monitoring". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0169-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Renal Div
- Product
- Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A.
- Product code
- KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Reason for recall
- The Meridian pump is less likely to detect small pressure changes, which may indicate the presence of a post-pump tubing kink, when using pre-pump arterial pressure monitoring.
- FDA root cause
- Other
- Action
- Urgent Device Correction letters dated 9/6/06 were sent to the Meridian accounts on the same date via UPS second day, to the attention of the Hemodialysis Administrator. The letters informed the accounts of the potential risk associated with a post-pump bloodline kink when using pre-pump arterial pressure monitoring. The accounts were provided with a revised Meridian Operator''s Manual which includes the following warnings: 'The lower venous limit should be set as close to the currently displayed venous pressure as conditions allow.' and 'Reliance on the pressure alarms should not substitute for clinical monitoring of the patient for adverse events.' Any technical questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 3, option 2, choice 2. General questions were directed to The Center for One Baxter at 1-800-422-9837.
- Quantity in commerce
- 2,823 units
- Distribution
- Nationwide and Hong Kong, Korea and China
- Product codes and lots
- all serial numbers