Z-0193-2007
BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging System
Trionix Research Laboratory, Inc. / initiated 2006-08-01 / Terminated
Trionix Research Laboratory Inc began a recall of BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging System on . The reason FDA records is "The ball screw on the unit which supports the detector head (which weighs approximately 800 lbs.) could snap resulting in the detector head dropping on the patient". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0193-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Trionix Research Laboratory, Inc.
- Product
- BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging System.
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- The ball screw on the unit which supports the detector head (which weighs approximately 800 lbs.) could snap resulting in the detector head dropping on the patient.
- FDA root cause
- Other
- Action
- The recalling firm modified one of the devices in 2002. The remaining device users were telephoned on/about 11/02/2006. To date the recalling firm has not taken any corrective action regarding devices with serial numbers B-102, B-103, B-104 or B-106 thru B-184.
- Quantity in commerce
- 76 devices
- Distribution
- Worldwide, including USA, Australia, Egypt, UK, Belgium.
- Product codes and lots
- Serial numbers B-102 through B-184