Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0322-2007

Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required)...

Terumo Cardiovascular Systems Corporation / initiated 2004-10-12 / Terminated

Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763 on . The reason FDA records is "Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0322-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.
FDA root cause
Other
Action
U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.
Quantity in commerce
4,740 for all affected products
Distribution
Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Product codes and lots
Serial numbers 0013, 0017, 0019 through 0039, 0041 through 0163, 0165 through 0200 and 0203 through 0244.