Cytomics FC 500 Flow Cytometry System with CXP Software Versions 2.0 & 2.1

Siemens Healthcare Diagnostics Inc. / initiated 2006-07-26 / Terminated / root cause: software

Beckman Coulter Inc began a recall of Cytomics FC 500 Flow Cytometry System with CXP Software Versions 2.0 & 2.1 on . The reason FDA records is "Mis-identification-If a panel or protocol is added to an existing worklist but the tube location is not specified, the CXP Acquisition software will run the last specified tube through the remaining protocols and will...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0197-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Cytomics FC 500 Flow Cytometry System with CXP Software Versions 2.0 & 2.1
Product code
GKZ Counter, Differential Cell
Reason for recall
Mis-identification-If a panel or protocol is added to an existing worklist but the tube location is not specified, the CXP Acquisition software will run the last specified tube through the remaining protocols and will generate results and printouts of these runs leading to a Mis-Identification condition.
Root cause tag
software
FDA root cause
Other
Action
Customer letter -- A Product Corrective Action (PCA) letter was sent the week of July 24, 2006 to all accounts that have FC500 with CXP Software 2.0 and 2.1 to inform them that a potential sample misidentification may occur when running a Work list with missing tube Location(s) in Automatic MCL mode. The letter outlines actions to be taken to avoid mis-identification of patient results.
Quantity in commerce
416 (US) + 60 (Canada)
Distribution
Nationwide and Canada
Product codes and lots
Part Numbers: 722645 Software Kit; CXP version 2.1 New User; 723113 Software Kit; CXP version 2.1 Upgrade; 629637 Software Kit; CXP version 2.0 New User; 629636 Software Kit; CXP version 2.0 Upgrade; 626553 TN; FC500 WITH UPS; 6605627 TN; 5 CLR, FC500 (IOOV); 6605628 TN; 5 CLR, FC500 (120V); 6605629 TN; 5 CLR, FC500 (220V); 6605630 TN;5 CLR, FC500 (240V)