Z-0332-2007
Terumo Advanced Perfusion System 1 Temperature Monitoring System; Catalog number 802114
Terumo Cardiovascular Systems Corporation / initiated 2004-10-12 / Terminated
Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1 Temperature Monitoring System; Catalog number 802114 on . The reason FDA records is "Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0332-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Advanced Perfusion System 1 Temperature Monitoring System; Catalog number 802114.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.
- FDA root cause
- Other
- Action
- U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.
- Quantity in commerce
- 4,740 for all affected products
- Distribution
- Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
- Product codes and lots
- Serial numbers 00019 through 00034, 00036, 00037, 00046 through 00053, 00055 through 00069, 00073 through 00114, 00116 through 00128, 00131 through 00193, 00195 through 00203, 00205 through 00334, 00339 through 00355, 00357 through 00414, 00416 through 00431. 00434, 00435, 00439 through 00447, 00448 through 00466, 00468 through 00473, 00475 through 00477, 00479 through 00582, 00584 through 00590, 00592 through 00594 and 00606.