Z-0333-2007
Terumo Advanced Perfusion System 1 Flow Module; Catalog number 802018
Terumo Cardiovascular Systems Corporation / initiated 2004-10-12 / Terminated
Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1 Flow Module; Catalog number 802018 on . The reason FDA records is "Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0333-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Advanced Perfusion System 1 Flow Module; Catalog number 802018.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.
- FDA root cause
- Other
- Action
- U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.
- Quantity in commerce
- 4740 for all affected products
- Distribution
- Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
- Product codes and lots
- Serial numbers 00017, 00020, 00023 through 00026, 00029 through 00031, 00033, 00034, 00036 through 00048, 00051 through 00055, 00057 through 00078, 00080 through 00120, 00122 through 00134, 00136, 00138 through 00155, 00158 through 00169, 00171 through 00210, 00212 through 00243 and 00248.