Z-0058-2007
Convertors Tiburon Cardiovascular Split Drape II, Sterile, for single use only; Cardinal Health, McGaw Park...
Medtronic, Inc. / initiated 2006-09-27 / Terminated
Cardinal Health began a recall of Convertors Tiburon Cardiovascular Split Drape II, Sterile, for single use only; Cardinal Health, McGaw Park, IL 60085-6787, Made in Mexico The cardiovascular... on . The reason FDA records is "The cardiovascular drape may tear and/or fray at the reinforced fenestrated trough area during use". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0058-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Convertors Tiburon Cardiovascular Split Drape II, Sterile, for single use only; Cardinal Health, McGaw Park, IL 60085-6787, Made in Mexico The cardiovascular drapes were packagaged under the following configurations: a) Catalog #9158 - 1 CV split drape b) Catalog #9158CE - 8 x 1 CV split drape
- Product code
- KKX Drape, Surgical, Antimicrobial
- Reason for recall
- The cardiovascular drape may tear and/or fray at the reinforced fenestrated trough area during use.
- FDA root cause
- Other
- Action
- Cardinal Health sent separate recall letters dated 9/27/06 via UPS next day delivery to those accounts who ordered the drapes alone and those accounts who ordered custom sterile packs and PBDS modules, informing them that the cardiovascular split drapes may tear and/or fray at the reinforced fenestrated trough area during use. The letters listed the catalog numbers and affected lot numbers of the sterile drapes, and the letters sent to the pack and module customers were accompanied by a listing of catalog numbers and work order numbers for the affected packs ordered by each account. The accounts were requested to inspect their inventory to determine if they had any of the affected product on hand; segregate the affected inventories; complete and return via fax at 847-689-9101, the enclosed acknowledgment form, indicating the number of drapes in inventory being destroyed; and contact the El Paso Customer Service Center at 915-779-3681 to order replacement product. Distributors were requested to notify their customers of the recall. An additional letter dated 10/25/06 was sent via UPS overnight delivery to 49 of the accounts, expanding the recall to include lot 06HRD110 of catalog #9158, Convertors Tiburon Cardiovascular Split Drape II, Sterile. The instructions remained the same as those in the 9/27/06 letter.
- Quantity in commerce
- 1086 units
- Distribution
- Worldwide Distribution- USA and countries of Kuwait and Australia
- Product codes and lots
- Catalog #9158: lot numbers 06FRD326, 06FRD341, 06GRD006, 06GRD013 and 06HRD110; Catalog #9158CE: lot numbers 06FAD344, 06FAD370, 06FAD405, 06HAD109