Asahi APS Series Multiple-Use Hollow Fiber Dialyzers; a wet model multiple use dialyzer consisting of hollow...
Asahi Medical Co., Ltd began a recall of Asahi APS Series Multiple-Use Hollow Fiber Dialyzers; a wet model multiple use dialyzer consisting of hollow fiber membranes of polysulfone housed within a... on . The reason FDA records is "There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple Use Dialyzers at the end of treatment, possibly damaging the dialyzer nozzle". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0171-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Asahi Kasei Medical MT Corp. Oita Works
- Product
- Asahi APS Series Multiple-Use Hollow Fiber Dialyzers; a wet model multiple use dialyzer consisting of hollow fiber membranes of polysulfone housed within a plastic casing of styrene butadiene block polymer, filled at the factory with fluid to facilitate priming by the user, and sterilized by gamma radiation before shipment; Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Japan; Made in Japan; 12 units per case; Models APS-18R and APS-900S - 1.8 m2 surface area, Model APS-21R and APS-1050S - 2.1 m2 surface area
- Product code
- KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Reason for recall
- There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple Use Dialyzers at the end of treatment, possibly damaging the dialyzer nozzle.
- FDA root cause
- Other
- Action
- Asahi sent Safety Alert letters on 10/12/06 to their re-use dialyzer customers advising them of the reports of difficulty separating the blood set from the dialyzer and not to use the Gambro Cartridge Blood Set with Asahi Reuse/Multiple Use Dialyzers in a reuse situation. Enclosed were copies of the temporary revision Instructions For Use with the following warning added at the top of the page: 'Additional Warning for Reuse/Multiple Use Dialyzers: Ashai Reuse/Multiple use dialyzers must not be used in connection with the Gambro cartridge Blood Set in a reuse environment.' Any questions were directed to Karen Borling, RN at 847-498-8500.
- Quantity in commerce
- 100,000 for all models
- Distribution
- Nationwide
- Product codes and lots
- Model APS-18R, Model APS-900S, Model APS-21R and APS-1050S, all lots